next up previous contents
Next: 4.1.1 CCD Subassembly Calibration Up: 4 Calibration/Maintenance Previous: 4 Calibration/Maintenance

4.1 Lab Data

The laboratory calibration of ACIS produces models based on the measurements of:

Calibration is done by illuminating the CCDs with low energy X-rays from a monochromator or hard X-rays from a radioactive source. All CCDs have been tested, but some were tested more than others (deeper exposures, for example, which are too time-consuming to perform on all CCDs, but which can be used to characterize them). Absolute efficiency measurements have been done at the BESSY synchrotron radiation facility in Berlin on Lincoln Lab CCDs made in the same way as the flight CCDs. These chips provide transfer standards for assessing the absolute quantum efficiency of all the CCD chips flown in both the imaging and spectroscopic arrays, and the flight backup chips.



 
next up previous contents
Next: 4.1.1 CCD Subassembly Calibration Up: 4 Calibration/Maintenance Previous: 4 Calibration/Maintenance

John Nousek
11/21/1997