Next: 4.1.1 CCD Subassembly Calibration
Up: 4 Calibration/Maintenance
Previous: 4 Calibration/Maintenance
The laboratory calibration of ACIS produces models based on the
measurements of:
-
quantum efficiency, energy resolution, and system
gain and linearity as a function of incident energy and instrument
parameters (such as temperature and clock voltages);
-
non-X-ray light rejection efficiency, including UV/visual optical
blocking filter checks;
-
CCD read-noise, charge transfer efficiency, charge collection
efficiency, and dark current as a function of energy and CCD parameters;
-
geometric layout of CCDs relative to ACIS fiducial marks,
including seam locations and sizes;
-
limits to six charge transfer clock levels -- high and low on each of: imaging,
frame-store, and serial units -- for optimum CTE, noise, and dark current;
-
relationship between pre-amplifier bias, reset bias, reset clock
levels and read noise, amplifier responsivity, and reset
performance (the time to clear the CCD, or dead time);
-
verification of read-out modes.
Calibration is done by illuminating the CCDs with low energy X-rays
from a monochromator or hard X-rays from a radioactive source. All
CCDs have been tested, but some were tested more than others (deeper
exposures, for example, which are too time-consuming to perform on all
CCDs, but which can be used to characterize them).
Absolute efficiency measurements have been done at the BESSY
synchrotron radiation facility in Berlin on Lincoln Lab
CCDs made in the same way as the flight CCDs. These chips
provide transfer standards for assessing the absolute quantum
efficiency of all the CCD chips flown in both the imaging
and spectroscopic arrays, and the flight backup chips.
Next: 4.1.1 CCD Subassembly Calibration
Up: 4 Calibration/Maintenance
Previous: 4 Calibration/Maintenance
John Nousek
11/21/1997