4.1.2.2 Assumptions for spectral resolution measurements

1.
Spectral resolution is effectively independent of position on the CCD.
2.
Quantum efficiency measurements provide spectral resolution data at energies different from those specified here (see next section).


  
Table 4.1: Spectral Resolution Measurements
System Energy Total Cts No. of
  (keV) (/energy/CCD quad.) Frames
HIREFS Spec. 0.1-1.5 $5\times 10^{5}$ per band 5000
       
HEXS 1.489 (Al-K) $6 \times 10^{5}$ per energy $400\times 4$
  2.622 (Cl-K)    
  6.404 (Fe-K)    
  9.886 (Ge-K)    
       
HEXS 4.952 (V-K) $6 \times 10^{5}$ per energy $400\times 4$
  6.930 (Co-K)    
  7.478 (Ni-K)    
  8.639 (Zn-K)    
       
Si Edge Spec. 1.0-2.2 $5\times 10^{5}$ per band 5000